Running the Built-in Diagnostics and
Checking the Installed Options
The instructions below allow you to run the built-in diagnostic
program in your HP8920A, HP8920B, HP8921A and 8924E/C units. This will
test your unit for most faults. When you first turn on your unit, the
screen
should indicate:"All self tests pass". This message indicates that the
power supply voltages are correct and there are no major failures. It
is not an indication that the unit is working properly. Therefore, you
should run the built-in diagnostics for a better system evaluation.
To run your unit’s self diagnostics including: Audio Frequency, Radio
Frequency and Misc. performance (you will need (2) BNC-BNC cables, (1)
BNC "T" adapter and (1) N male to BNC female adapter):
1)Press the “TESTS” button
2) Select “ROM” under "select
procedure location". (select by moving the cursor with the large scroll
knob / select by pressing in the scroll knob)
3) Select “AF_DIAGS”; or
“RF_DIAGS”; “MISC_DIAGS(run only part 2- port tests)” "under
procedure filename". If you are testing an 8924E/C unit, select CDMA
diags also. Depending on the firmware and model, the diagnostic
procedure may be named AFDIAG2; RFDIAG2; MSDIAG2, etc.
4) Select RUN.
5) Follow on-screen directions.
You will need to attach the BNC cables to different ports. Each test
may take from 2 - 6 minutes. At end of test, screen will indicate # of
failures. Note any failures. These procedures extensively test the
HP892x product. There is a possibility that the unit will pass all
these self diagnostics and still have a fault, but this is rare.
To check your installed options:
1) Press the “TESTS” button.
2) Select “ROM” under procedure
location.
3) Select LIST_OPTS under procedure
4) Select RUN
The HP8920 will now list the included options on the screen.